You've already made the smart choice by choosing our logger, now make the most of it!
The below instructions help you get started with your logger. If you have any further questions or concerns, please don't hesitate to contact us. Have a great day!
To begin, we recommend you first create an account at log.temperature.cloud. This account serves as your personalized portal to register your logger(s) and base station(s).
One of the best parts?
One account allows you to register and manage an unlimited number of loggers and base stations.
To receive updates, tips and tricks, instruction videos, and more in your mailbox!
🔋 First things first, charge the logger's battery for 6 hours.
🔌 Connect a MICRO-USB cable to the logger and a charging device.
⏱️ Enjoy up to 6 weeks of uninterrupted usage at 20°C with the fully loaded built-in rechargeable lithium-ion battery.
Driver is only compatible with Windows.
🔌 For first-time computer connection, install the driver:
1️⃣ Download the software (link)
2️⃣ Check the video
3️⃣ Once installed, you're good to connect!
We recommend installing our driver on your computer to:
Set the time of your logger
Upload your temperature measurements manually. (For example in case you don’t have a Base Station, or you have no data plan for your Base Station)
Configure your Base Station
(Only applicable to first use)
1️⃣ Connect logger to computer using MICRO-USB B (not included).
2️⃣ Activate USB mode: press and hold "Start" and "Pause" buttons for 3 seconds; LED light turns orange (logger won't record).
3️⃣ Open browser, type "http://CC-L-A-”logger serial no” (for example, for logger 00000121, type "http://cc-l-a-00000121/".
4️⃣ Go to Advanced Settings, Press “Set Now and Save” or configure date and time manually. Press Save and disconnect.
5️⃣ After disconnecting the logger, the LED turns red (recording restarts after 5 seconds)
Start measurement: press the Start button for 3 seconds
Quick green LED flashes confirm the activation
During recording the green LED will flash shortly once a minute
Pause measurement: press the Pause button for 3 seconds
Quick red LED flashes confirm the pause mode
As long as the logger is in pause mode, the red LED will flash once a minute
Stop measurement: press the Pause button for more than 8 seconds
Quick red LED flashes will appear, hold the Pause button until the LED will turn off
No more LED flashing, no more measurement
Logger is not detectable by any Base Station
This is the “hibernation” mode (for long term storage)
USB mode: press and hold Start and Pause buttons for 3 seconds
LED light turns orange
No measurement
Enter flight mode: press and hold Start for 5 seconds
LED flashes orange
No radio transmission
Exit flight mode: press and hold Start for 3 seconds
LED flashes green
Radio transmission on
1️⃣ Connect logger to computer using MICRO-USB B (not included).
2️⃣ Activate USB mode
3️⃣ Open browser, type "http://CC-L-A-”logger serial no” (for example, for logger 00000121, type "http://cc-l-a-00000121/".
4️⃣ Press download - the file will be stored on your PC.💾
5️⃣ After disconnecting the logger, the LED turns red (recording restarts after 5 seconds)
1️⃣ Go to https://log.temperature.cloud/upload/
2️⃣ Click "Uploading", select your file. Let the data flow 📈!
3️⃣ Login to visualize & analyze.
Nijverheidsstraat 8 | 2235 Hulshout | Belgium | +32 15 67 67 68